Daily Shutdown =============== These procedures are used for shutting down MagAO-X at the end of a work day. Minimal shutdown ---------------- This procedure can be used if you expect to startup in a few hours, right away the next day, etc. The main goal here is to shutoff "the expensive stuff". Namely: - dmtweeter - dmwoofer - dmncpc - camwfs - camsci1 - camsci2 - camlowfs - ttmmod - ttmpupil Rules: - do not power off a DM unless it has been released - do not power off a camera unless its temp is 20C or higher - do not power off instcool! - steps can be done in parrallel, so start warming up cameras, go on, then come back Cameras (each of camsci1 camsci2 camlowfs camwfs): - close the shutter - warm up to 20C (use cameraGUI, change setpoint) - wait for it to finish - power off - note: for camlowfs, power off under both pdu1 and usbdu0. dmtweeter: - open the loop - press "loop zero" - zero flat - zero all ch - release - power off dmwoofer and dmncpc - zero flat - zero all ch - release - power off ttmmod: - if modulating, first press set in Pupil Alignment GUI - press rest - power off in Power GUI ttmpupil - press rest in Pupil Alignment GUI - power off Additionally: - power off camtip (its just a heater) - power off source - power off tableair Remaining devices such as focus stages and filter wheels can be left on, which will make alignment easier next time. Standard Daily Shutdown ----------------------- If MagAO-X will be unused for a longer period of time (say a weekend), we typically shut down everything but computers, networking and cooling. At the end of this, the following things will still be on: - pdu0.compicc - pdu0.comprtc - pdu0.dcpwr - pdu0.swinst - pdu3.blower - pdu3.fanaux - pdu3.fanmain - pdu3.instcool **Note:** it is critical that you not shutdown instcool while compicc and/or comprtc are on! Follow the steps in "Minimal Shutdown" above. Also do the following. fwtelsim and fwscind: - first power off on usbdu0 - then power off on dcdu1 Everything else (except the items listed above) can be shutdown in any order. Occassional errors in the logs during shutdown may occur.