Polarimetry
Point of contact: Miles Lucas <mileslucas@arizona.edu>
Warning
This document is a work in progress. If you plan to do polarimetry, please email me beforehand.
The procedures below use the Polarimetry GUI, but every step can also be done with INDI properties directly. If the GUI isn’t running, launch it from exao1 with
nohup polarimetryGUI &
Setup
Align the instrument following the standard alignment procedure. Leave the fwscind pupil lens in and the J-test on; you will need them to realign the pupil after inserting the QWPs.
Prepare devices
Power on the following stages using
pwrGUI:vpdu1.stagepolrotvpdu1.stageqwprotvpdu0.stagezaber(in the ninja tab)
Insert the HWP
Open the AO loop (
holoop).Open the pupil tracking loop in the Pupil Alignment GUI.
In the Polarimetry GUI, set HWP lin. stage to
in.
Insert the QWPs
Make sure the AO loop and the pupil tracking loop are both open.
If they aren’t there already, set
stageqwprot1andstageqwprot2to thelabpreset usingcursesINDI.In the Polarimetry GUI, set QWP lin. stage to
in.Realign the system pupil:
Danger
Do not use Auto Alignment here. It is not calibrated for the pupil position after a QWP insertion.
Roughly realign the pupil using the J-test and the ttmperi arrow buttons in the Pupil Alignment GUI.
Close the AO loop (
holoop).Watch the Pupil Tracking Loop offsets and adjust ttmperi until both are within \(\pm 0.05\).
The pupil alignment loop offsets in the Alignment GUI
In the Polarimetry GUI, enable QWP pupil tracking. This lets
ttmpericorrect the pupil shifts as the QWPs rotate.Enable QWP pol. tracking. This lets the QWPs stabilize and optimize the eigenpolarizations of the instrument.
Beamsplitter, filters, and cameras
Set
stagescibstopol.Put
fwsci1andfwsci2in the same filter (r, i, or z).Important
camsci1 and camsci2 must use the same filter. This is PDI, not SDI.
Enable Synchro in either camsci control GUI, then tune the exposure time and EM gain for both cameras.
Setup is complete. You can now power off stagezaber in the ninja tab of pwrGUI.
Sequencing
Set the sequencing parameters in the Polarimetry GUI:
Num. cycles: the number of HWP cycles to run. Set to 0 to run until stopped.
Frames per pos.: the number of camera frames at each HWP position. This is meant for long exposures and requires an exposure time of at least 1 s. Set to 0 to use a timer instead (the default).
Time per pos.: the time spent at each HWP position, in seconds. Only used when Frames per pos. is 0.
In the Observers interface of the MagAO-X Web UI, fill in the observer, operator, target, and observation names.
Click Start sequencing. To stop:
Last cycle: finish the current HWP cycle, then stop.
Stop sequencing: finish the current HWP position, then stop.
Abort: stop immediately.
Timing and efficiency
A HWP cycle is one full pass of imaging at 0°, 45°, 22.5°, and 67.5°. Each cycle carries 19 ± 2 s of overhead from the HWP moves and camera triggering, regardless of the time per position.
Time per pos. (s) |
Cycle (s) |
Overheads |
Observing efficiency |
|---|---|---|---|
5 |
39 |
49% |
51% |
10 |
59 |
33% |
67% |
15 |
79 |
24% |
76% |
30 |
139 |
14% |
86% |
60 |
259 |
7% |
93% |
The total time for a sequence can be estimated as
Teardown
Remove the QWPs
Open the AO loop (
holoop).In the Polarimetry GUI, disable QWP pol. tracking.
Disable QWP pupil tracking.
Open the pupil tracking loop in the Pupil Alignment GUI.
In the Polarimetry GUI, set QWP lin. stage to
out.
Remove the HWP
Make sure the AO loop and the pupil tracking loop are both open.
In the Polarimetry GUI, set HWP lin. stage to
out.Roughly realign the pupil using the J-test and ttmperi.
Close the AO loop (
holoop).In the Pupil Alignment GUI, run Auto Alignment until both pupil offsets are within \(\pm 0.05\), then turn it off.
Close the pupil tracking loop.
Details
Polarizing beamsplitter (PBS)
The polarizing beamsplitter cube (PBS) splits the incident light into orthogonal linear polarizations, one for each science camera: camsci1 measures horizontally polarized light and camsci2 measures vertically polarized light. Polarimetric observables come from differencing the two cameras’ images or fluxes,
and they are normalized by the sum,
To use the PBS, set stagescibs to pol, either through INDI or from either camsciCtrl GUI. In the data, check the STAGESCIBS PRESET NAME FITS header; it should read pol.
Half-wave plate (HWP)
Because the PBS has a fixed orientation, the cameras alone can only measure horizontal and vertical polarization (Stokes Q). To measure +45° and −45° polarization (Stokes U), the half-wave plate (HWP) rotates the incoming polarization so that Stokes U lands on the horizontal/vertical axes of the PBS.
The HWP is also critical for cancelling instrumental polarization. In the same way it rotates Stokes U onto Q, it can flip incident +Q to −Q. Two measurements at different HWP angles, one measuring +Q and one measuring −Q, let us cancel the instrumental term differentially:
To use the HWP, insert it into the beam with stagepollin. The HWP is controlled by the hwptrack application. To move it to a given angle, set hwptrack.hwp_position.target. The standard sequence for linear Stokes polarimetry steps through 0°, 45°, 22.5°, and 67.5°, which measure Stokes +Q, −Q, +U, and −U, respectively. The current position name is reported in hwptrack.hwp_position_name.value and can be used to group frames during data analysis.
The following FITS headers are provided for data analysis:
HWPTRACK SET ANGLE / [deg] Commanded fast-axis angle of the HWP, typically 0, 45, 22.5, or 67.5
HWPTRACK ACTUAL ANGLE / [deg] Actual fast-axis angle of the HWP; differs from SET ANGLE only when tracking is on
HWPTRACK POS NAME / Named HWP position: 'Qplus', 'Qminus', 'Uplus', 'Uminus', or 'unknown'
HWPTRACK TRACKING / [int] 1 if synchro_adi tracking is on, 0 if off
Dual rotating quarter-wave plate compensator (DQWP)
The DQWP can reorient the eigenpolarizations of the instrument. In practice, this means it keeps light that enters horizontally or vertically polarized horizontally or vertically polarized when it is measured. Its main use is to compensate for the cross-talk between polarization states caused by the k-mirror image rotator.
The following FITS headers are provided for data analysis:
QWPTRACK QWP1 ANGLE / [deg] Fast-axis angle of QWP1
QWPTRACK QWP2 ANGLE / [deg] Fast-axis angle of QWP2
QWPTRACK TRACKING / [int] 1 if QWP pol. tracking is on, 0 if off